4.3 Transducer for Non-Electrical Biosignals
|
113
4.3.2 Optical Sensors for Plethysmography and Determination of Oxygen
Saturation
Plethysmography (Gr. plethys = fullness) means the recording of volume variations
within the body. In the context of this book, we restrict ourselves to the volume change
in vessels due to pulsatile blood flow. The palpation of the pulses belongs to the oldest
diagnostic procedures and plays a major role especially in traditional Chinese medi-
cine. On surface (subcutaneous) vessels, the change in volume can also be measured
optically. In this case it is called photoplethysmography. The measuring principle is
that light is irradiated into the body and the transmitted or reflected radiation is meas-
ured with a light sensor. The sensor signal is variable in time as a result of the blood
flow. When the pulse wave passes through the irradiated vessel section, the vessel
dilates due to the larger blood volume. The larger blood volume then ensures greater
light absorption compared to the phase when there is only a small volume of blood
in the vessel section. As a result, the light sensor detects less light. In this way, con-
tinuous pulse curves can be recorded. The pulse curve contains information about
the condition of the vessels and the cardiovascular system. For a differentiated eval-
uation, pulse curves usually have to be measured at several points. It is particularly
easy to determine the heart rate with photoplethysmography. To do this, the duration
between the maxima of the pulse curve is determined. The reciprocal of this multiplied
by 60 gives the heart rate in the usual unit of beats per minute.
An extension of photoplethysmography is the SpO2 measurement⁸. It allows the
determination of the oxygen saturation of the blood. For this purpose, the different ab-
sorption of hemoglobin (Hb) and hemoglobin saturated with oxygen (oxyhemoglobin
HbO2) at different wavelengths is exploited. For the determination of oxygen satura-
tion, measurements at at least two different light wavelengths are necessary, so at least
two monochromatic light sources are required. The measurement result is expressed
as a percentage and corresponds to the ratio of HbO2 to total hemoglobin (HbO2 + Hb).
The procedure is presented in more detail in subsection 6.3.2.
Semiconductor light-emitting diodes (LEDs) are predominantly used as the light
source. These have a high intensity with comparatively low power consumption, have
a small design and are very inexpensive. When measuring SpO2 with two LEDs, care
must be taken to ensure that the emission spectra are far enough apart and do not
overlap. LEDs are available in the color spectrum from blue to infrared. Color selection
is based on the absorption spectra of Hb and HbO2. In order to distinguish between
the transmitted and reflected radiation of the respective LEDs at the detector, the two
LEDs must be operated in pulse mode and multiplexing mode. The accuracy of the
SpO2 determination depends crucially on the adherence to the emission wavelength,
as stated in subsection 6.3.2. From Table 4.3, it can be seen that the location of the
8 S denotes saturation, p pulse, O2 oxygen.